|8/31/2013 | |8/31/2013 | ||
|Ryan Pierce | |Ryan Pierce | ||
|Installed network software in C:\NETWORK. | |Installed network software in C:\NETWORK. Used EZSETUP.EXE to configure Ethernet card for 0x280, IRQ 3, 0xCC00 as recommended in Leica | ||
manual. (Note: Memory address conflicts on factory setting. Use soft config jumper.) Set up DOS packet driver. Set up for BOOTP. Networking | |||
under DOS works, and the DHCP server will respond with the appropriate static IP. Moved unnecessary startup items to a new group so they do not launch on startup. | |||
|- | |- | ||
|8/30/2013 | |8/30/2013 | ||
|Ryan Pierce | |Ryan Pierce | ||
|Changed from backup hard drive to system hard drive. Removed laser printer card. Installed SMC Elite 16 Ethernet card. Set up template | |Changed from backup hard drive to system hard drive. (Backup that Jay made works. Note that he must have taken the backup when the SEM | ||
user directory in C:\USER\TEMPLATE and set up C:\NEWUSER.BAT script to create new users. Put sensible default settings in the Template | was in a weird state; you will need to load one of the configs in the default user directory just to see beam and an image. The LUT may have the colors inverted too. Reminder: have Jay make another backup once we're good to go with networking.) Removed laser printer card. Installed SMC Elite 16 Ethernet card. Set up template user directory in C:\USER\TEMPLATE and set up C:\NEWUSER.BAT script to create new | ||
dir. | users. Put sensible default settings in the Template dir. | ||
|- | |- | ||
|7/23/2013 | |7/23/2013 | ||