Rockwell Hardness Tester: Difference between revisions
From Pumping Station One
| Line 49: | Line 49: | ||
=== Procedure === | === Procedure === | ||
1. Select the penetrator and test block | 1. Select the penetrator and test block | ||
| Line 64: | Line 56: | ||
2. Set up the tester | 2. Set up the tester | ||
Secure the penetrator and the appropriate anvil in the tester | Secure the penetrator and the appropriate anvil in the tester. | ||
3. Apply Minor Load | 3. Apply Minor Load | ||
Slowly turn the hand wheel to bring the indicator hand to the line | Slowly turn the hand wheel to bring the small dial indicator hand to the line. This applies minor pressure load to the penetrator. | ||
4. Apply Major Load | |||
Turn the hand wheel only until the dial pointer rests on the major pressure load indicated by the chart. | Turn the hand wheel only until the dial pointer rests on the major pressure load indicated by the chart. | ||
| Line 82: | Line 71: | ||
Turn the hand wheel back to bring the indicator hand back to “SET” and take the reading on the barrel dial through the Lucite magnifier. Each graduation on the barrel dial indicates two points in the Rockwell scale. | Turn the hand wheel back to bring the indicator hand back to “SET” and take the reading on the barrel dial through the Lucite magnifier. Each graduation on the barrel dial indicates two points in the Rockwell scale. | ||
5. Take the reading | |||
Standard Tester: | Standard Tester: | ||
| Line 89: | Line 78: | ||
With the ball penetrator, read column B (red numbers). | With the ball penetrator, read column B (red numbers). | ||
NOTE: The first two or three tests may be low until the penetrator and anvil have become firmly seated. | NOTE: The first two or three tests may be low until the penetrator and anvil have become firmly seated. | ||