Leica S430 Scanning Electron Microscope: Difference between revisions
From Pumping Station One
| Line 232: | Line 232: | ||
! Maintainer | ! Maintainer | ||
! Description | ! Description | ||
|- | |||
|4/29/2014 | |||
|Ryan Pierce | |||
|Disassembled Penning gauge, straightened anode wire better, cleaned O ring with kimwipe, and reassembled. System pressure is lower, measured at 1.09e-4 torr on 5/1/2014. | |||
|- | |||
|4/26-27/2014 | |||
|Ryan Pierce and Elizabeth Koprucki | |||
|Disassembled Penning gauge. It was filthy with plated material and loose shiny black sand-like material. Sanded anode and cathode cups to remove the bulk of plated material. Removed ion trap and sanded interior metal surfaces. (The plastic surfaces are filthy, but I don't know how to clean this.) Cleaned all metal parts in 3 progressive changes of acetone in an ultrasonic bath. Anode wire was bent in re-insertion. Following re-assembly, pressure was 1.5e-4 torr. | |||
|- | |- | ||
|4/7/2014 | |4/7/2014 | ||