|Ryan Pierce | |||
|Discovered that tension on the HRRU fiber cables snapped the HRRU fiber connector of one cable and damaged the locking mechanism on OPTO 11, which likely caused the loss of vacuum and loss of communication to the vacuum board. Removed the HRRU cables so they are no longer a hazard. Re-seated all fiber connections to the image processor board. | |||
|- | |- | ||