Technics Hummer V Sputter Coater: Difference between revisions
From Pumping Station One
RyanPierce (talk | contribs) No edit summary |
No edit summary |
||
| Line 12: | Line 12: | ||
|where = Across from Woodshop | |where = Across from Woodshop | ||
|value = $5,000.00 | |value = $5,000.00 | ||
|image = Hummer_V_spatter_coater.jpg | |||
}} | }} | ||
Revision as of 01:04, 30 September 2013
| This tool is currently broken, please contact the area host for more information. |
Configuration
- Standard sputter coater
- Carbon arc attachment for carbon deposition
Status
- Requires argon for operation. Preferably high purity argon.
- Power cord is scary and needs to be replaced.
- Gas flow rate meter has a small leak, needs a new gasket.
- Changing the vacuum oil would be a good idea.
- Initial tests with welding argon showed that it worked and was sputtering metal, however maintaining a plasma seemed to use a lot of current. The former operator suggested pulsing it to burn any oxygen in the chamber out. The manufacturer suggested a short that could be remedied by a good cleaning, which was performed.
List of Currently Certified Users
| Qualified Member | Trained By |
|---|---|
| Ryan Pierce | Ryan Pierce |
| Steve Finkelman | Steve Finkelman |
Documentation
Manual scans for the sputter coater are available, as is a manual for a similar but not identical model of the carbon arc deposition system. A location to drop them is needed.
The manufacturer was purchased by Anatech USA and they still support this product.
Maintenance Log
| Date | Maintainer | Description |
|---|---|---|
| 7/21/2013 | Ryan Pierce | Tried using this after cleaning. Vacuum could only get down to 30-40 millitorr. The power supply seemed to need excessive amperage just to maintain a plasma, and that was with low voltage. Then something inside the power supply exploded with a burning smell. |
| 4/?/2013 | Steve Finkelman | Full cleaning of chamber. |