|5/1/2014
|Ryan Pierce
|Disassembled Penning gauge, straightened anode wire better, noted some of the particulates on the O ring, cleaned it better with a kimwipe, and reassembled. System pressure is now 9.3e-5 torr, which is sufficient to run the beam.
|-
|-
|4/29/2014
|4/29/2014