|5/1/2014 | |||
|Ryan Pierce | |||
|Disassembled Penning gauge, straightened anode wire better, noted some of the particulates on the O ring, cleaned it better with a kimwipe, and reassembled. System pressure is now 9.3e-5 torr, which is sufficient to run the beam. | |||
|- | |- | ||