| Line 212: | Line 212: | ||
On a single occasion, communication to the EHT board was lost. This corresponds to Opto 8 and 9. | On a single occasion, communication to the EHT board was lost. This corresponds to Opto 8 and 9. | ||
A dual cable plugs into a single port, Opto 1, for the HRRU pixel data. The fiber connector | A dual cable plugs into a single port, Opto 1, for the HRRU pixel data. The fiber plug connector on the cable end has snapped. This is not critical as we are not using the HRRU. | ||
Repair is | Repair is possible. The connector in question is a "hp 9534 R-2521" with a T-2521 next to it. These appear commonly available, but soldering to the image processor board should be considered a last resort. | ||